Reference:
Documenting Process Calibrator LR-Cal LPC 300. Measurement of Current (0…20 mA, 4…20 mA, resolution 1 µA, accuracy ±1.6 µA). Measurement of Voltage (0…1/2/5/10 V, resolution 0.1 mV, accuracy ±0.5 mV). Sourcing 24 VDC as power supply for test items (e.g. for pressure transmitter etc.). Conversion current/voltage to pressure. Operated with chargeable Lithium-Ion battery, battery charger with EU/US/UK-adapters included, large display. Programmable resolution, up to 6 digits. Simultaneous display of test pressure, value of test item, deviation and deviation as percentage of full scale value of test item. Operating modes: MEASURE, CALIBRATE and SWITCH-TEST. Creation (storage) of calibration procedures, storage of calibration results. Changeable reference-sensors LPC-S (plug & play, up to 5 LPC-S reference sensors can be handled with one LPC 300 unit, this means, up to 5 different pressure ranges with one instrument). Programmable display resolution, up to 6 digits. Switchable pressure units: bar/mbar, psi, inHg, cmHg, mH2O, cmH2O, mmH2O, kg/cm², mmHg plus 1 user-definable. Functions: Min/Max values (positive and negative PEAK function), Tare, Min-/Max-Alarm, Filter, Zerro adjustment, Powersave. With USB- and RS232 inferface. Electrical connections: 4 mm sockets for measurement of current/voltage, 24 VDC output, switch test. The LPC-S reference sensors are supplied with pressure connection 1/2″ BSP male, suitable for the reference port of the LR-Cal LPP calibration handpumps. Incl. 1.1 m cable to connect LPC-S reference sensor with LPC 300 pressure calibrator.
PC-Windows Software LR-Cal LPC-Cal and USB-cable included. LPC-Cal allows online-calibrations and printing of certificates. (Operating system MS Windows 98 SE, 2000 SP4 or XP neccessary; installed MS Excel version 2000, XP or 2003 necessary).
LR-Cal LPP-KIT-PK-LPC300 ±0.025% f.s.
Available LPC-S pressure ranges:0…0.25 bar, 0…0.4 bar, 0…0.6 bar, 0…1 bar, 0…1.6 bar, 0…2.5 bar, 0…4 bar, 0…6 bar, 0…10 bar, 0…16 bar, 0…25 bar, 0…40 bar gauge pressure
-0.25…0 bar, -0.4…0 bar, -0.6…0 bar, -1…0 bar, -0.25…+0.25 bar, -0.4…+0.4 bar, -0.6…+0.6 bar, -1…+1.5 bar, -1…+3 bar, -1…+5 bar, -1…+9 bar, -1…+15 bar, -1…+20 bar, -1…+24 bar, -1…+39 bar
0…0.4 bar absolute, 0…0.6 bar absolute, 0…1 bar absolute, 0…1.6 bar absolute, 0…2.5 bar absolute, 0…4 bar absolute, 0…6 bar absolute, 0…10 bar absolute, 0…16 bar absolute, 0…25 bar absolute pressure
Content:
- 1 calibration test pump LR-Cal LPP 30
- 1 test hose 0.5 m with pressure port 1/4″ BSP female rotating for test item
- 1 set of BSP thread adapters
- 1 reference instrument LR-Cal LPC 300 Electronic Pressure Calibrator
- 1 battery charger with EU/US/UK adapters
- 1 reference sensor LPC-S
- 1 electr. cable (1.1 m) to connect LPC-S sensor to LPC 300 unit
- 1 USB cable
- 1 CD-ROM with software LR-Cal LPC-Cal (incl. USB driver)
- 1 traceable certificate of calibration for current and voltage
- 1 traceable certificate of calibration for pressure
- Operating Manuals
- 1 aluminum transit case with custom foams
LR-Cal LPP-KIT-HK-LPC300 ±0,025% f.s.
Available LPC-S pressure ranges:
0…100 bar, 0…160 bar, 0…250 bar, 0…400 bar, 0…600 bar, 0…700 bar, 0…1000 bar
Content:
- 1 calibration test pump LR-Cal LPP 700 or LR-Cal LPP 1000 (depending on range)
- 1 test hose 1.0 m with pressure port 1/4″ BSP female rotating for test item
- 1 set of BSP thread adapters
- 1 reference instrument LR-Cal LPC 300 Electronic Pressure Calibrator
- 1 battery charger with EU/US/UK adapters
- 1 reference sensor LPC-S
- 1 electr. cable (1.1 m) to connect LPC-S sensor to LPC 300 unit
- 1 USB cable
- 1 CD-ROM with software LR-Cal LPC-Cal (incl. USB driver)
- 1 traceable certificate of calibration for voltage and current
- 1 traceable certificate of calibration for pressure
- Operating Manuals
- 1 aluminum transit case with custom foams
LR-Cal LPP-KIT-PH-LPC300 ±0,025% f.s.
Combined kit, available pressure ranges: see above.
Content:
- 1 calibration test pump LR-Cal LPP 30
- 1 test hose 0.5 m with pressure port 1/4″ BSP female rotating for test item
- 1 calibration test pump LR-Cal LPP 700 or LR-Cal LPP 1000 (depending on range)
- 1 test hose 1.0 m with pressure port 1/4″ BSP female rotating for test item
- 1 set of BSP thread adapters
- 1 reference instrument LR-Cal LPC 300 Electronic Pressure Calibrator
- 1 battery charger with EU/US/UK adapters
- 2 reference sensors LPC-S
- 1 electr. cable (1.1 m) to connect LPC-S sensor to LPC 300 unit
- 1 USB cable
- 1 CD-ROM with software LR-Cal LPC-Cal (incl. USB driver)
- 1 traceable certificate of calibration for voltage and current
- 2 traceable certificates of calibration for pressure
- Operating Manuals
- 1 aluminum transit case with custom foams
Email This Page